5-lead ECG cable | 1pc |
Surface temp.probe | 1pc |
Adult NIBP cuff | 1pc |
NIBP tube | 1pc |
Adult SpO2 sensor | 1pc |
Power cord | 1pc |
Electrode | 5pcs |
User manual | 1pc |
1. Standard parameters:ECG,RESP,NIBP,SPO2,PR/HR,TEMP.
2. 13 types Arrhythmic Analysis, Multi-lead ECG waveforms display in Phase, 15 types Drug calculation and titration table.
3. Anti-ESU, anti-defibrillator; RA-LL impedance.
4. SPO2 can testing for 0.1% weak,high accuracy.
5. Unique breathing measurement technology, real-time monitoring, intelligent filter design, eliminate interference, measuring respiratory accurately.
6. Trend Coexist interface for all parameter to observe easily.
7. OxyCRG Dynamic View display, More intultive with all parameters.
8. Platform scalable with cable or wireless connecting to central monitoring.
9. Printer /Recorder is optional for real-time printing.
10. Brief operation interface, Touchscreen is optional. Up to 2-5 hours working capacity of built-in rechargeable Li-ion battery.
11. 12.1"high resolution color TFT LCD display.
12. Holographic waveform trends playback for 2 weeks and 12 hours off electricity saving.
13. Built-in networking to realize central monitoring function.
14. Optional: Touchscreen, VGA, 2-IBP, HDMI, ETCO2.
15. System start up only need 10 seconds with ARM-A5 Nuclear technology.
1. ECG Lead Mode | 5 Leads or 3 Leads |
3 Leads | RA, LA, LL, Lead mode: I, II, III |
5 Leads | RA, LA, LL, RL, V, Lead mode: I, II, III, aVR, aVL, aVF, V, Gain x250, x500, x1000, x2000 |
Heart Rate |
|
Range | Adult, 15 ~ 300bpm; Neonate / Pediatric, 15 ~ 350 bpm |
Accuracy | ±1%or ±1bpm, |
Resolution | 1 bpm |
Sensitivity | >200 uV |
Differential Input Impedance | >5MΩ |
Bandwidth |
|
Diagnostic | 0.05~130Hz |
Monitor | 0.5~40Hz |
Surgery | 1~20Hz |
Method | Impedance between R-F(RA-LL) |
Scope of base impedance | 200~4000Ω |
Bandwidth | 0.1~ 2.5 Hz |
RESP | |
Range | Adul, 2~120BrPM; Neonate / Pediatric, 2~150 BrPM |
Resolution | 1 BrPM |
Accuracy | ±2 BrPM |
Suffocation warning | 10~40s |
Method | Oscillometric |
Mode | Manual, Auto, STAT |
Measuring interval in AUTO MODE | 1,2,3,4,5,10,15,30,60,90,120,180,240,480 MINS |
Measuring interval in STAT MODE | 5 MINS |
PR range | 40~240 bpm |
Measuring Range and Resolution | |
Adult | SYS, 40~270mmHg; DIA, 10~215mmHg; MEAN, 20~235mmHg |
PED | SYS, 40~200mmHg; DIA, 10~150mmHg; MEAN, 20~165mmHg |
NEO | SYS, 40~135mmHg; DIA, 10~100mmHg; MEAN, 20~110mm; HgSTATIC NIBP range, 0~300mmHg |
Static NIBP accuracy | ±3mmHg |
Max Average tolerance ±5mmHg | Max. Standard tolerance; 8mmHg |
Over-pressure protection | |
Adult mode | 300 mmHg |
Ped mode | 240 mmHg |
Neo mode | 150 mmHg |
Measuring range | 0~100% |
|
|
Resolution | 1% |
|
|
Accuracy | 70~100% ±2 | 0%~69% | Not defined |
Pulse Rate |
|
|
|
Measuring range | 20~300bpm |
|
|
Resolution | 1bpm |
|
|
Accuracy | ±3bpm |
|
Channels | 2 |
Measuring |
|
Range | 0~50°C |
Resolution | 0.1°C |
Accuracy | ±0.1°C (sensor tolerance no not included) |
Adapter |
|
Input | AC 90 V~240V, 50/60Hz |
Output | DC 12V-20V |
Built-in Battery |
|
TYPE | 14.8V, 2200mAH Li-ion rechargeable battery |
Operating time | 3.5-4 hours working time under full power |
Charging time | 2 hours |
Label | ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
Measuring and alarm range |
|
ART | 0 ~ 300 mmHg, PA -6 ~ 120 mmHg |
CVP/RAP/LAP/ICP | -10 ~ 40 mmHg, P1/P2 -10 ~ 300 mmHg Press Sensor |
Sensitivity | 5 uV/V/mmHg, Impedance 300-3000Ω |
Resolution | 1 mmHg |
Accuracy | ±2% or ±1mmHg, which great |
Actualization interval | about 1 Sec. |
Measuring and alarm range | 0% ~ 13% |
Resolution | 1 mmHg |
Accuracy | 2 mmHg @ < 5.0% CO₂ (at ATPS) |
RR | 3 - 150 bpm |